DUROCELL - Accessories

Additional stainless steel or chrome wire mesh screen (protective Epolon layer)
Additional mesh screen for the unit chamber. Allows processing of a larger amount of material. See the summary of technical parameters for the maximum recommended number of pieces depending on the inner chamber volume.
.

Drain vat – sheet metal – stainless steel (protective Epolon layer)
(not available for 22-L volume)
Serves for the condensate collection.
.

Printer
Allows archiving of the process course in the unit chamber. A heading showing the unit type, preset temperature and/or the ramp, number of cycles, fan speed, and the preset print time interval is printed on the paper tape. The following values are then printed on one line below the heading: Time from the program start and actual temperature in the chamber (and/or the flexible sensor temperature is printed below it). The printer is connected to the printing socket of the unit by means of a cable with a serial connector (RS 232C). The printer is connected to the electric power mains by an adapter. A heading showing the unit type, preset temperature and/or the ramp, number of cycles, fan speed, RH, light, and the preset print time interval is printed on the printer. The following values are then printed on one line below the heading: Time from the program start and actual temperature in the chamber (and/or the flexible sensor temperature is printed below it).
.

3-point measurement incl. IQ and OQ reporting
9-point measurement incl. IQ and OQ reporting
The compliance of the unit temperature course must be of maximum accuracy and reproducibility. Records confirming the compliance with the quality standards like ISO 9000 and ISO/IEC 17025 are the basic condition for proving the quality of services provided by certified laboratories.

Installation qualification (IQ)
IQ proves and evidences that the equipment and supporting systems have been chosen adequately, installed properly, and comply with the required specifications.

Operational qualification (OQ)
OQ proves and evidences that the unit can perform the proposed functions repeatedly and reliably within the extent of the operational parameters.
.

Chip card (in Comfort-line only)
The programs installed in the unit may subsequently be modified at any time by the unit user by means of a chip cards system. The chip cards serve for an unlimited storing and recording of the user programs.
.

Movable unit table (except for 22-L, 404-L, and 707-L volume)
The table forms one body with the unit. Wheels enable easy moving.
.