« back

FRIOCELL - Optional equipment

(please specify at ordering)

access ports Ø 25, 50,100 mm (Ø 100 mm not available in 22-L volume)
The diameter, number, and location of the access ports is optional. The metal access ports are closed from the outer side by a special plastic plug enabling cables passing through from the outer space into the chamber.
.

 

Door lockdoor lock
Mechanical lock is situated on the upper front surface of the door near the closing mechanism. It serves for the unit opening prevention by an unauthorized person.
.

Zrcadlové dveřeleft hinged door (except for 22, 404 and 707-L volume)
The door is a model with mirror symmetry to the right hinged door.
.

separate PT100 sensor
An additional freely movable sensor inside the chamber, by means of which the temperature inside the chamber or in the material can be measured. The sensor cables are connected to a connector located in the rear part of the unit.
.

WarmCommspecial software WarmComm 4 B
The software is intended for recording the temperature regulation course in the temperature cabinets. The data acquired during the regulation are displayed in form of a graph (where the horizontal and vertical axis shows the time and acquired data, respectively). The program allows regulation monitoring in real time, saving the regulation course to a file on a disc, and viewing the formerly created files. The WarmComm 4 Basic software allows data monitoring of more than 10 connected units.
.

komunikační software WarmCommspecial software WarmComm 4 Professional or WarmComm 4 F
The software is intended for recording the temperature regulation course in the temperature cabinets. The data acquired during the regulation are displayed in form of a graph (where the horizontal and vertical axis shows the time and acquired data, respectively). The program allows the regulation monitoring in real time, saving the regulation course to a file on a disc, and viewing the formerly created files.
- The extended version of the WarmComm 4 Professional software allows monitoring of the unit data and controlling the unit.
- The full version of the WarmComm 4 F software meets the requirements of FDA 21 CFR Part 11. Optional cable length: 5, 10, and 15 m.
.

stainless steel unit jacket AISI 304
High design quality and long service life of the unit
.

time-programmable inner waterproof socket (4 pins)
Allows connection of suitable equipment (shaking machine, mixer) placed in the unit chamber.

BMS relay alarm contact
Allows remote control of the unit operation. It is connected to a connector in the rear leg of the unit (24 V/1A).
.

cooling from as low as -9.9oC
In case lower working temperature than 0°C is required, we offer a model enabling working temperature from as low as -9.9 °C
.

inner lighting
wide range of various lighting sources (except for 22-L volume)

exposure /stimulating lighting
(white/day light) with digital setting of intensity from 10 – 100 %; door 6,000 – 13,000 Lx, shelves – 12,000 – 16,000 Lx (depending on the volume)

exposure lighting in shelves
Especially for the photostability testing (according to ICH Guideline CPNP/ICH279/95)
.

light exposure measurement
Wide range of various light sources offer (except for 22-L volume)
UV – ultraviolet radiation - PC/printer output
VIS – visible radiation - PC/printer output
.

open-door control (microswitch)
Opening of the door during a program run is recorded on the connected printer or in the WarmComm software record in the connected PC.
.

automatic door locking
Prevents access of unauthorized persons
.