« back

STERIVAP - Optional equipment

(must be specified in the order))

one-door and two-door (pass-through) version

.

optional steam source
FD - steam from a central source
ED - integrated steam generator
FED - combined mode of supply by technical/medicinal steam
.

stainless steel panel sheets of the device and covering strips
optional
.

integration in the stainless steel partition walls
available at the two-door version (type 336-2)
.

polishing of the inner surface of the sterilization chamber
with roughness Ra 0.8 µm (Ra 32 µinch), Ra 0.125 mm (Ra 5 μinch)
.

system for manual loading of materials
1) basic insert for containers and baskets
2) basic insert and frame (for max. 2 sieves)
3) basic insert, frame, shelves/sieves guide, shelves or sieves (max. 5 pieces)
4) suitable for laboratories – basic insert, solutions tank, solutions frame (for 1 sieve) – for sterilization of solutions

.

shelves guides

.

wire sieve

.

perforated shelf

.

transport and loading carts system
insert for the loading cart, transport cart, cassette-type loading cart, container-type loading cart, shelf-type loading cart, loading cart for solutions sterilization
.

frame for the loading cart

.

transport cart

.

cassette-type loading cart

.

container-type loading cart

.

shelf-type loading cart

.

loading cart for solutions sterilization

.

dripping vat for solutions in the sterilization chamber

.

hook for the loading carts unloading

.

sterilization basket
570 × 280 × 135 mm (D × W × H) - 1/2 STI
575 × 280 × 265 mm (D × W × H) - 1 STI
.

sterilization container

.

additional mechanical manometers
- on the loading side
- on the unloading side
.

optional electrical connection depending on the required network parameters

.

stainless steel device framework

.

stainless steel pipe interconnection
- FD version; ED / FED version
.

Rostfreies Sicherungsventil
.

stainless steel valves and fittings
.

stainless steel safety valve
.

flexible PT 100 sensor for sterilization of solutions

.

special programmes
- Sterilization of solutions with spontaneous jacket cooling and supporting air pressure (contains also the flexible PT100 temperature sensor; requires the Ekom plus 2V compressor)
- Solutions sterilization process controlled by Fo values
.

laboratory software(to change the temperature and time of sterilization)
allows the operator to make individual adjustments in the pre-programmed sterilization programmes.
.

special UNICONFIG software
allows modification of the individual phases of the sterilization cycle (evacuation, vacuum depth, exposure, drying) and to set the temperature and time values (verification with the manufacturer is required)..
.

pecial Printer Archiv software
to document the batches in a PC
.

Software Ecosoft
to connect the sterilizer to the LAN network
.

special programmes on chip cards
.

emergency door sealing
in case of power failure; without a compressor
.

emergency door sealing
to check continuously the content of air and incondensable gases in the sterilizing chamber during each sterilization programme, thus achieving maximum sterilization safety as compared to routine controls by means of testing programmes (vacuum and Bowie&Dick test) that are carried out only once daily before commencement of common operation (HTM 2010).
.

„Automatischer Morgen-Start“
is another of the series of cost saving products that will save your time. The device switches on in absence of the operator according to the pre-set time. The device is pre-heated automatically and the Vacuum Test is performed.

.

version with low-energy cost-saving steam generator
Optimierter Leistungsbedarf 8 kW
.

tropical version
for countries with high temperature of cooling water
.

version with autonomous operation
eliminates the need of the device connection to a source of water. The device is equipped with an integrated cooling aggregate which is placed in the rear part of the device.
.

version with an integrated steam/steam exchanger
allows production of medicinal steam intended for sterilization, without any need of an integrated electric steam generator and its supply by a powerful electrical source. Technical steam serves as a heat source in this case. Technical steam is available very often in health care facilities. As a consequence, this kind of steam production represents lower running costs and cost saving thanks to the lower demands on electrical distribution system rating.
.

Energetic Maximum function
set of software and hardware adjustments of the sterilizer that enable individual and mutual regulation of integrated steam generators to ensure control of maximum energy take-off and communication with the technical centre, or that enable limitation of power distribution rating at the installation site.
.

thermo-degassing of the supply demi-water
minimizes the content of incondensable gases in the steam generator. Integrated equipment for condensate cooling is available.
.

media monitoring
continuous control of input media parameters (pressure air, demi-water and cooling water)
.

3 F socket
(3-phase socket for appliances)
.

water treatment equipment
(for ED / FED models only)
.

air compressor
- including the air accumulator and cabinet DK 50 S
- including the air accumulator and cabinet Ekom plus 2V
.

monitoring packet of indicators

.

testing and validation according to EN 285+A2 and EN ISO 17665-1

.

basic documentation for IQ / OQ / PQ

.

FAT, SAT takeover

.

8 GB memory card
for the sterilization cycles recording (up to 100,000 hours of record)
.

"Audit trail"
- system events recording on the memory card (conforms to 21CFR part 11)
.