« back

STERIVAP HP IL - Optional equipment

(must be specified in the order)

one-door and two-door (pass-through) version
Stainless steel panel sheets; they may be built in stainless steel partition walls; mirror device version enables – if more devices are installed one next to another – joining of two service areas in a one.
.

děrovaná policeoptional steam source
FD - steam from a central source
ED - own integrated steam generator
FDT - combined mode of supply by technical/medicinal steam
FDD - integrated own steam/steam exchanger
.

komunikační software WarmCommpolishing of the inner surface of the sterilization chamber 
Roughness Ra 1,25 µm (Ra 50 µinch); Ra 0.8 µm (Ra 32 µinch); Ra 0.125 µm (Ra 5 µinch)
.

rám pro vozíkroughness Ra 1,25 µm (Ra 50 µinch); Ra 0.8 µm (Ra 32 µinch); Ra 0.125 µm (Ra 5 µinch)
Chamber rails for transfer of the loading cart. The rails are made of solid square tube for high resistance against damage and for increase of the load capacity. The rails are equipped with a mechanism for the transport cart fixing. The rails are made of stainless steel, type DIN 1.4404 (AISI 316L).
.

vedení policsystem for manual material loading – sieve and shelf supports
Shelf holder + 2 stainless steel perforated shelves. The sterilization chamber is equipped with a special frame fixed to the inner side of the chamber. It is made of stainless steel, material no. 1.4404 (AISI 316L). There are holes punched in the frame to fix the shelf supports. The upper shelve can be placed up to 9 different positions. The shelves are made of stainless steel, material no. 1.4301 (AISI 304), and there are holes punched in them, enabling easy circulation of steam.
.

stainless steel valves with screwed or welded necks of the CLAMP type

.

stainless steel safety valve

.

gas-tight Bio-Seal version of the device 
with the possibility of independent and permanent sealing of the chamber door by pressurized air and independent door opening at any side
.

special stainless steel, sterilisable filters on the sterilization chamber input and output
- Bacteriological filter on the chamber output (decontamination including the condensate sterilization
- Bacteriological sterilisable air-inlet filter on the air supply side with preparation for the integrity test
.

PT 100 temperature sensor
.

special programmes on chip cards 
Chip cards system + 10 cards. The Chip Card function is activated and appears in the Main Menu of the sterilizer. The Chip Card system enables saving of sterilization programmes from the chip cards to the selected positions in the sterilizer, using all free positions (up to 14 programmes). The function enables also saving of sterilization programmes from the sterilizer to the chip card – one chip card can bear up to 2 sterilization programmes. The function enables direct copying of the sterilization programmes from the already occupied positions to the free programme positions (up to 14 programmes). 
.

dripping vat for solutions
A vat which prevents solutions leakage. It is a special vat for collection of solutions leaked out of bottles during the sterilization cycle. It is made of stainless steel, type DIN 1.4301 (AISI 304).
.

condensate cooling
Condensate cooling equipment may be built in. The condensate is cooled in connection with a water ring vacuum pump. Thermostatically controlled cooling equipment instead of the standard condensing equipment of the device. 
.

"Air Detector"
The function of the Air Detector is to check continuously the content of air and incondensable gases in the sterilizing chamber during each sterilization programme, thus achieving maximum sterilization safety as compared to routine controls by means of testing programmes (vacuum and Bowie&Dick test) that are carried out only once daily before commencement of common operation (HTM 2010). Air detector: Software function and hardware adjustment of the sterilizer for continuous measurement of the content of air in the sterilization chamber. In case the measured content of air in the chamber exceeds the set limits, the process is automatically stopped and a relevant error message appears.
.

additional mechanical manometers 
- on the loading side
- on the unloading side
.

dripping vat under the device

.

large touch screen 12" even on the unloading side

.

bar code reader 

.

special Printer Archive software for batches documentation in a PC 

.

software for the sterilizer connection to the computer network (LAN) 

.

chamber passivation (acid cleaning) 

.

laboratory software
It enables the operator to perform individual adjustments in the already programmed programms.
.

special programmes
Sterilization of solutions with spontaneous cooling and supporting air pressure (contains also the flexible PT100 temperature sensor), Solutions sterilization process controlled by Fo values. Special MOVEX®/UNICONFIG software enables modification of the individual phases of the sterilization cycle (evacuation, vacuum depth, exposure, and drying) and setting of the sterilization cycle temperature and time (verification with the manufacturer is required).
.

media monitoring
Continuous checking of input media parameters (pressurised air, demi-water, and cooling water).
.

energetic Maximum function
Regulation of the device operation. Control of maximum energy take-off in case of more devices connection into the electric network.
.

tropical version for countries with high temperature of cooling water

.

optional electrical connection depending on the requested mains parameters 
.

transport cart
Transport cart for the loading cart. The cart has a solid chassis, transport plate, pushing handle, and four wheels. The front wheels are lockable in the direction of movement; the rear wheels can be locked by parking brakes. Coupling system for the BMT loading cart connecting and disconnecting. The cart construction, wheels and brakes are made of stainless steel. The wheels are equipped with soft tyres.
.

komunikační software WarmCommloading cart – for solutions

.

Zrcadlové dveřeloading cart – universal
Loading cart for transportation of items to be sterilized to the sterilization chamber and for the chamber unloading. The cart has four wheels and construction with four shelf holders. The frame and the moving gear are made of stainless steel, material DIN 1.4301 (AISI 304). The cart has a bottom perforated shelf and an additional perforated shelf in the upper fixed position. The edges are folded upwards to secure the loaded goods.
.

loading cart - special

.

děrovaná policestainless steel shelf 

.

stainless steel sieve (except for 446 and 636) 

.

hook for the loading carts unloading

.

large scale of laboratory accessories
Bags and sacs for contaminated material, sterilization baskets, plastic vessels, test tubes, Petri dishes, etc.
.

basic IQ, OQ, and PQ documentation for validation according to GMP and GLP
.

tests and validations according to EN 285:06 and EN ISO 17665-1:06 

.

air compressor including the air accumulator and cabinet (for devices with an additional Solution Programme with Forced Jacket Cooling and Supporting Air Pressure, a more powerful air compressor is required , e.g. Ekom plus 2 V) 

.

water treatment equipment for demi-water preparation 

.

starting packet of monitoring indicators 

.

optional language version for communication with the device 

.

 
 
.

- 8 GB memory card for the sterilization cycles recording (up to 100,000 hours of record).  

.

Automatic sterilizer door opening during a power failure  

.

"Audit trail" - system events recording on the memory card (conforms to 21CFR part 11) 
 
.